beautypg.com

Measuring mode, alarms, Alarm indications – RKI Instruments GX-2012 Manual User Manual

Page 44

background image

36 • Measuring Mode, Alarms

GX-2012 Operator’s Manual

Measuring Mode, Alarms

This section covers alarm indications in Measuring Mode. It also tells you
how to reset the GX-2012 after an alarm has occurred and how to respond to
an alarm condition.

NOTE: False alarms may be caused by radio frequency (RF) or

electromagnetic (EMI) interference. Keep the GX-2012 away from
RF and EMI sources such as radio transmitters or large motors.

Alarm Indications

CAUTION: If the combustible channel is set for %volume only, then there

are no gas alarms for any detection channel. See “HC Range
Screen” on page 44.

The GX-2012 will sound an alarm, the unit will vibrate, and the LED arrays
will flash when one of the target gas concentrations rises above the Warning
level, or in the case of oxygen, falls below the Low Alarm setting for that gas.
The GX-2012 also sounds an alarm, vibrates, and flashes the LED arrays
when the Alarm level is reached for combustible gas, CO, and H

2

S, when the

concentration of oxygen rises above the High Alarm level, and when the
STEL and TWA alarm points are reached for CO and H

2

S.

When a failure condition occurs, such as a sensor failure, low flow, or dead
battery condition, the unit will also sound an alarm, flash the LED arrays, and
vibrate.
The table below summarizes the types of alarms produced by the
GX-2012.

Table 4: Alarm Types and Indications

Alarm Type

Visual Indications

Other Indications

Low Alarm

Concentration of gas

rises above the Warning

level, or falls below the

Low Alarm level for O

2

.

• WARNING appears

below the gas list.

• Reading for the gas in

alarm flashes

• Alarm LED arrays

flash

• Backlight turns on

• Buzzer sounds

alternating between a

low and high pitch

• Vibrator pulses