beautypg.com

Specifications – RKI Instruments GX-2012 Manual User Manual

Page 11

background image

GX-2012 Operator’s Manual

Specifications • 3

Specifications

Table 1: GX-2012 Specifications

Target Gas

%LEL

Combustible

Gas

(Methane

Calibration

Standard)*

% Volume

Combustible

Gas

(Methane

Calibration

Standard)

Oxygen (O

2

)

Hydrogen

Sulfide

(H

2

S)

Carbon

Monoxide

(CO)

Range

(Increment)

0-100% LEL

(1% LEL)

0 - 100% vol

(1% vol)

0-40.0% vol

(0.1 vol%)

0-100 ppm

(0.5 ppm)

0-500 ppm

(1 ppm)

Leak Check

Mode

Range

(Increment)

0-5,000 ppm

(100 ppm)

NOTE: Detection
range is 0-5,000
ppm, but displayed
range can be set to
0-500 ppm, 0-1000
ppm, 0-2000 ppm,
or 0-5000 ppm.

N/A

N/A

N/A

N/A

Sampling

Method

Sample Draw

Response

Time

T90 Within 30 Seconds

Accuracy

± 5% of reading

or ± 2% LEL

(whichever is

greater)

± 5% of

reading or

± 2% full

scale

(whichever

is greater)

± 0.5% O

2

± 5% of

reading or

± 2 ppm H

2

S

(whichever

is greater)

± 5% of

reading or

± 5 ppm

CO

(whichever

is greater)

Display

Digital LCD Display

Gas Alarms

(Factory

Settings)

Alarm 1

10% LEL

Alarm 2

50% LEL

None

Alarm 1

19.5 vol%

(Decreasing)

Alarm 2

23.5 vol%

(Increasing)

Alarm 1

5 ppm

Alarm 2

30 ppm

TWA

1 ppm

STEL

5 ppm

Alarm 1

25 ppm

Alarm 2

50 ppm

TWA

25 ppm

STEL

200 ppm

Operating

Temperature

& Humidity

-20°C to 50

°

C/Below 85% RH (Without Condensation)