GX-2012 Operator’s Manual

Using Maintenance Mode • 103

5. Once you enter Maintenance Mode, MAINTENANCE will appear along

the top of the screen. The Date Screen displays.

6. Use the AIRS or (SHIFT)T button to move through the menu item

screens.

Updating the Date and Time Settings

1. Use the AIRS or (SHIFT)T buttons to display the DATE menu item.

Press and release the POWER ENTER button. The date and time will be
displayed with the year flashing.

2. Use the AIRS or (SHIFT)T button to display the desired year.
3. Press and release the POWER ENTER button to save the setting. The

month setting flashes.

4. Repeat steps 2 and 3 to enter the month, day, hours, and minutes settings.

The Date menu item displays after you enter the minutes setting.

Performing a Fresh Air Adjustment

1. Confirm you are in a fresh air environment that is free of toxic and

combustible gases and of normal oxygen concentration, 20.9%.




MAINTENANCE

DATE




MAINTENANCE

DATE

2011
4
23
22:37

MAINTENANCE

DATE