Dlog dsp, Bump dsp, Turning the bump test function on or off – RKI Instruments GX-2012 Manual User Manual

Page 133

background image

GX-2012 Operator’s Manual

Using Maintenance Mode • 125

set to OFF, the Clear Data Logger Screen does not appear in Display Mode
and any logged data cannot be cleared from Display Mode.
1. Use the AIRS or (SHIFT)T buttons to display the DLOG DSP menu

item.

2. Press and release the POWER ENTER button. The screen will indicate

the current setting.

3. Use the AIRS or (SHIFT)T button to display the desired setting.
4. Press and release the POWER ENTER button to save the setting and

return to the main menu.

Turning the Bump Test Function On or Off

The BUMP DSP menu item is used to turn the bump test function on or off.
If set to OFF, the BUMP menu item in Calibration Mode and Maintenance
Mode does not appear. In addition, the BUMP--SET, BP INTVL, BP
RMNDR,
and BP EXPRD menu items will not display in Maintenance
Mode. The factory setting is OFF.
1. Use the AIRS or (SHIFT)T buttons to display the BUMP DSP menu

item.


MAINTENANCE

DLOG DSP

On


MAINTENANCE

DLOG DSP


MAINTENANCE

BUMP DSP