86 • Start Up, Leak Check Mode

GX-2012 Operator’s Manual

NOTE: If no button is pressed for 20 seconds, the unit will proceed into

whichever mode is displayed.

6. If CL RMNDR is turned on (factory setting), the screen that appears next

depends on how CL EXPRD is set in the Maintenance Mode Menu
(page 131).

If the unit is due for calibration and CL EXPRD is set to CONFIRM
(factory setting), then the following screen displays.

The alarm LED’s and buzzer will pulse several times. After this,
press and release the RESET SILENCE button to continue.

If the unit is due for calibration and CL EXPRD is set to NOT USE,
then the following screen displays.

The GX-2012 cannot be used until a calibration has been performed
either by selecting AUTO CAL, or ONE CAL in the Maintenance
Mode menu. See “Calibrating Using Auto Calibration” on page 105
or “Calibrating Using Single Calibration” on page 108 for calibration
instructions.