92 • Leak Testing

GX-2012 Operator’s Manual

2. Move the probe tip or tapered nozzle tip back and forth along the area

where a leak is suspected.

3. Observe the display reading. If the gas level increases, the numerical

reading will increase, the bar graph level will increase, and the beeping
and buzzer pulsing frequency will increase.

4. Your gas reading may exceed your display range or 100% LEL. For a

description of both scenarios, see the following section.

5. Use the increasing and decreasing of the reading to locate the leak point.
6. To exit Leak Check Mode and return to the Leak Check Mode Select

Screen at any time, press and hold the (SHIFT)T button for 5 seconds.

Overscale Conditions

There are three overscale conditions that the GX-2012 may experience while
operating in Leak Check Mode.
• Exceeding display range

If you are using a display range of 500, 1000, or 2000 ppm and the gas
level goes above the display range, the full scale value of the display
range will be displayed but it will not be an overscale condition unless the
gas level is above 5000 ppm.

• Exceeding 5000 ppm

If the combustible gas level exceeds 5000 ppm, the GX-2012 will display
an overscale condition but will continue displaying the gas reading in
%LEL.

32

5000

%LEL

ppm

LEAK CK MODE

CH

4

OVER