Agilent Technologies Agilent E5250A User Manual
Page 94

4-24
Agilent E5250A User’s Guide, Edition 11
Setting up Measurement Environment
Measurement Cable Length
Table 4-6
Guard Capacitances of E5250A Measurement Environment (Typical)
Connection
Agilent Model/Part No.
Cable
Length
Guard
Capacitance
from SMU to
E5250A
16494A-003 (for non-Kelvin)
80 cm
75 pF
16494B-003 (for Kelvin)
80 cm
90 pF
16494C-001 (for 4142B)
1.5 m
140 pF
E5250A and
Plug-in Card
E5252A (only 1 card is installed)
−
145 pF
E5252A (more 1 card)
−
8 pF/extra card
E5255A (only 1 card is installed)
−
180 pF
E5255A(more 1 card)
−
60 pF/extra card
from Card
Output to
Connector
Plate
16494A-001 (for non-Kelvin)
1.5 m
125 pF
16494A-002 (for non-Kelvin)
3 m
240 pF
16494B-001 (for Kelvin)
1.5 m
140 pF
16494B-002 (for Kelvin)
3 m
260 pF
16494C-001 (for 4142B)
1.5 m
140 pF
16494C-002 (for 4142B)
3 m
260 pF
16494D-001
1.5 m
125 pF
16494D-002
3 m
240 pF
16494E-003
3 m
240 pF
from
Connector
Plate to DUT
Agilent Part No. 8121-1191
X m
130 pF per m