General Specifications
General Specifications
Temperature
Range
Operating: 5
°C to 40 °C
Storage:
−40 °C to 70 °C
Humidity Range
Operating: 5 % to 80 % RH, non-condensing
Storage: 5 % to 90 % RH at 65
°C
Altitude
Operating: 0 to 2,000 m (6,500 ft)
Storage: 0 to 15,240 m (50,000 ft)
Regulatory
Compliance
Safety: CSA C22.2 No. 1010.1 / IEC61010-1
EMC: CISPR 11 Group 1 class A&EN50082-1
Power
Requirement
100 to 240V ± 10 %, 47 to 63 Hz, maximum 100 VA
Number of Slots
4 slots for 30-mm height switch cards
Dimensions
Mainframe: 230 mm H
× 430 mm W × 600 mm D
Option 001: 30 mm H
× 395 mm W × 424 mm D
Option 501: 30 mm H
× 395 mm W × 424 mm D
Weight
Mainframe: 11.2 kg
Option 001: 2.0 kg
Option 501: 2.1 kg
Number of Ports
with option 001
Low Leakage I-V Port: 2
General I-V Port: 4
C-V Port: 2
HF Port: 2
Output Channels: 12
Number of Ports
with option 501
Low Leakage I-V Port: 3
Bias Port: 3
Output Channels: 24 (8
Ч 3)