Option 501 multiplexer, Switch specifications – Agilent Technologies Agilent E5250A User Manual
Page 300

10-6
Agilent E5250A User’s Guide, Edition 11
Specifications
Option 501 Multiplexer
Option 501 Multiplexer
Switch Specifications
This specification is defined with the module plugged into the mainframe. The value
depends on the following temperature and humidity conditions.
•
A: 23
°C ± 5 °C, 5 % to 60 % RH
•
B: 5
°C to 18 °C, 28 °C to 40 °C, 5 % to 60 % RH
•
C: 28
°C to 40 °C, 60 % to 80 % RH
A B C
Max. Current Rating (A)
1.0
1.0
1.0
Max. Voltage Rating (V)
Channel to Guard
200
200
200
Channel to Channel
300
300
300
Closed Channel Residual Resistance (
Ω)
Low Leakage I-V Port
0.6
0.6
0.6
BIAS Port
a
1.0
1.0
1.0
Channel Isolation Resistance (
Ω)
Low Leakage I-V Port
10
13
5
Ч 10
12
2
Ч 10
12
BIAS Port
10
9
5
Ч 10
8
2
Ч 10
8
a. When 0
Ω resistors are mounted on the card as the protective resis-
tor.