Agilent E5250A User’s Guide, Edition 11

10-9

Specifications

Supplemental Data when used with Agilent 4155/4156

Supplemental Data when used with Agilent
4155/4156

The following data shows degradation of Agilent 4155/4156 when used with Low
Leakage I-V Port.

Iout is output current in A, and Vout is output voltage in V.

Supplemental Data when used with Agilent 4155

Voltage Range and Offset Accuracy (with MPSMU)

Current Range and Offset Accuracy (with MPSMU)

Supplemental Data when used with Agilent 4156 (non-Kelvin connection)

Voltage Range and Offset Accuracy (with HRSMU)

Current Range and Offset Accuracy (with HRSMU)

Range Setting

Measurement

2 V

960

μV + (0.5 × Iout) V

780

μV + (0.5 × Iout) V

Range Setting

Measurement

1 nA

3 pA + (0.2

× Vout) pA

3 pA + (0.2

× Vout) pA

10 nA

7 pA + (0.2

× Vout) pA

5 pA + (0.2

× Vout) pA

100 nA

50 pA + (0.4

× Vout) pA

30 pA + (0.4

× Vout) pA

Range Setting

Measurement

2 V

500

μV + (0.5 × Iout) V

280

μV + (0.5 × Iout) V

Range Setting

Measurement

10 pA

0.4 pA + (0.2

× Vout) pA

0.1 pA + (0.2

× Vout) pA

100 pA

0.4 pA + (0.2

× Vout) pA

0.1 pA + (0.2

× Vout) pA

1 nA

0.7 pA + (0.2

× Vout) pA

0.5 pA + (0.2

× Vout) pA

10 nA

4 pA + (0.2

× Vout) pA

2 pA + (0.2

× Vout) pA

100 nA

40 pA + (0.3

× Vout) pA

20 pA + (0.3

× Vout) pA