Supplemental data when used with agilent 4155/4156 – Agilent Technologies Agilent E5250A User Manual
Page 303
Agilent E5250A User’s Guide, Edition 11
10-9
Specifications
Supplemental Data when used with Agilent 4155/4156
Supplemental Data when used with Agilent
4155/4156
The following data shows degradation of Agilent 4155/4156 when used with Low
Leakage I-V Port.
Iout is output current in A, and Vout is output voltage in V.
Supplemental Data when used with Agilent 4155
•
Voltage Range and Offset Accuracy (with MPSMU)
•
Current Range and Offset Accuracy (with MPSMU)
Supplemental Data when used with Agilent 4156 (non-Kelvin connection)
•
Voltage Range and Offset Accuracy (with HRSMU)
•
Current Range and Offset Accuracy (with HRSMU)
Range Setting
Measurement
2 V
960
μV + (0.5 × Iout) V
780
μV + (0.5 × Iout) V
Range Setting
Measurement
1 nA
3 pA + (0.2
× Vout) pA
3 pA + (0.2
× Vout) pA
10 nA
7 pA + (0.2
× Vout) pA
5 pA + (0.2
× Vout) pA
100 nA
50 pA + (0.4
× Vout) pA
30 pA + (0.4
× Vout) pA
Range Setting
Measurement
2 V
500
μV + (0.5 × Iout) V
280
μV + (0.5 × Iout) V
Range Setting
Measurement
10 pA
0.4 pA + (0.2
× Vout) pA
0.1 pA + (0.2
× Vout) pA
100 pA
0.4 pA + (0.2
× Vout) pA
0.1 pA + (0.2
× Vout) pA
1 nA
0.7 pA + (0.2
× Vout) pA
0.5 pA + (0.2
× Vout) pA
10 nA
4 pA + (0.2
× Vout) pA
2 pA + (0.2
× Vout) pA
100 nA
40 pA + (0.3
× Vout) pA
20 pA + (0.3
× Vout) pA