Agilent E5250A User’s Guide, Edition 11

9-37

Executing Sample Programs

HCI Measurement Program

To change Idlin and Gmmax extraction condition

To change Gate voltage and Substrate voltage required for Idlin and Gmmax
extraction, do follows.

1. GET the PARAM.MES setup file by using the 4155/4156.

2. Change the value of Substrate voltage "VB" on the SOURCE SETUP screen.

3. Change the setting of "VG" parameter in "LINE1" on the ANALYSIS SETUP

screen.

To change the number of devices

Change the setting of the

No_of_devices

variable in the "Input Parameters" block.

To test more than 25 devices, you must change the configuration of the E5250A. See
chapter 2.

To change the stress times

Change the data at the DATA statement in the "Stress duration setup" block. Each
data shows the cumulative stress data.

Example:

DATA

10,

20,

50

DATA

100,

200,

500

DATA

1000,

2000,

5000

DATA

10000,

20000,

50000

DATA

100000

If you change the number of the data defined in DATA statements, change the value
of the

Meas_points

variable in the "Stress duration setup" block.

The value of

Meas_points

must be the same number as the number of data defined

in the DATA statements.

In the above example, the number of data in DATA statements is 13. So the variable
for the above example must be set as follows.

Meas_points=13