System:cconfig – Agilent Technologies Agilent E5250A User Manual
Page 211
Agilent E5250A User’s Guide, Edition 11
7-43
Command Reference
:SYSTem:CCONfig?
:SYSTem:CCONfig?
This query command returns the card configuration information. This command has
query form only.
Syntax
:SYSTem:CCONfig? card_number
Query response
card_configuration
For the E5252A: no response. Always returns #10.
For the E5255A: definite length block response data, #16ABCDEF, where each
letter is 8-bit binary data as follows:
A: indicates input port (1 to 6) that is connected to block 1.
B: indicates if resistance is set for block 1 Bias Port: 1: resistance set, 0: not set
C: indicates input port (1 to 6) that is connected to block 2.
D: indicates if resistance is set for block 2 Bias Port: 1: resistance set, 0: not set
E: indicates input port (1 to 6) that is connected to block 3.
F: indicates if resistance is set for block 3 Bias Port: 1: resistance set, 0: not set
Example
OUTPUT @Hp5250;":SYST:CCON? 1"
ENTER @Hp5250;A$
ENTER A$ USING "3X,6A";B$
FOR I=1 TO 6
C=NUM(B$[I;1])
PRINT "BYTE No. ";I;"= ";C
NEXT I
Parameter Explanation
card_number
card number: 1, 2, 3, or 4