beautypg.com

E5255a configuration examples, E5255a configuration examples -13 – Agilent Technologies Agilent E5250A User Manual

Page 45

background image

Agilent E5250A User’s Guide, Edition 11

2-13

Installation

Installing the E5250A

E5255A Configuration Examples

The following example configurations are described:

“24-output multiplexer (3-BIAS inputs, 1-IV input, 1 module)”

“24-output multiplexer (1-BIAS input, 1-IV input, 1 module)”

“96-output multiplexer (1-BIAS input, 1-IV input, 4 modules)”

24-output multiplexer (3-BIAS inputs, 1-IV input, 1 module)

To make this multiplexer, you need a E5255A. See Figure 2-3 for example setup.
The example uses the SMU INPUT 1 connector as the IV input, and uses 0

Ω

resistors.

Figure 2-3

24-output multiplexer (3-BIAS inputs, 1-IV input, 1 module)

1. Connect the IVin1 cable to the SMU1 input connector on the E5255A.

2. Connect the IVin2 cable to the IVout1 connector.

3. Connect the IVin3 cable to the IVout2 connector.