Route]:bias:port – Agilent Technologies Agilent E5250A User Manual
Page 195

Agilent E5250A User’s Guide, Edition 11
7-27
Command Reference
[:ROUTe]:BIAS:PORT
[:ROUTe]:BIAS:PORT
NOTE
This command applies only to the E5252A card.
For the E5255A card, the input Bias Port cannot be selected, and is always the
built-in Bias Port.
The Bias Port and Couple Port cannot be set to same port number.
This command selects which input port is the Bias Port on the specified card. For
each card, you can specify the same or different Bias Port. For details, refer to "Bias
Mode" in Chapter 5.
At *RST, Bias Port is set to 10 for the E5252A.
Syntax
[:ROUTe]:BIAS:PORT card_number,bias_port
Example
OUTPUT @Hp5250;":ROUT:BIAS:PORT ALL,6"
Parameter Explanation
card_number
card number:
For Auto Config mode: 0 or ALL
For Normal Config mode: 1, 2, 3, 4, or ALL
bias_port
Input port number: 1 to 10.