Programming the E5250A
Creating a Control Program
Table 6-1
Programming Summary
For details about these parameters, refer to Chapter 5 and Chapter 7.
CAUTION
Bias Port and Couple Port cannot be set for the same input port number.
Description
Command Default
Setting
1
Defines Channel Configuration Mode
:ROUT:FUNC channel_config
NCON
2
Defines Connection Rule
:ROUT:CONN:RULE card_no, rule
ALL,FREE
3
Defines Connection Sequence
:ROUT:CONN:SEQ card_no,sequence
ALL,BBM
4
Using Bias Mode
Selects input Bias Port (for E5252A)
:ROUT:BIAS:PORT card_no, input_port_no
ALL,10
Bias enables all output ports
:ROUT:BIAS:CHAN:ENAB:CARD card_no
ALL
Bias enables specified output ports
:ROUT:BIAS:CHAN:ENAB:LIST channel_list
All enabled
Bias disables all output ports
:ROUT:BIAS:CHAN:DISA:CARD card_no
All enabled
Bias disables specified output ports
:ROUT:BIAS:CHAN:DISA:LIST channel_list
All enabled
Sets Bias Mode ON/OFF
:ROUT:BIAS:STAT card_no, state
ALL,OFF
5
Using Couple Port (for E5252A)
Sets up Couple Port
:ROUT:COUP:PORT card_no, 'input_port_no'
No couple ports
Sets Couple Port Mode ON/OFF
:ROUT:COUP:STAT card_no, state
ALL,OFF
6
Controls Relays on Cards
:ROUT:CLOS channel_list
All relays open
:ROUT:OPEN channel_list
All relays open
:ROUT:OPEN:CARD card_no
ALL
Parameter Description
channel_config
NCON or ACON.
rule
FREE or SROU.
sequence
NSEQ, BBM or MBBR.
state
ON (1) or OFF (0).
channel_list
5 digit decimal number. See Chapter 5.
card_no
ALL or 0 or 1 to 4 are available.
input_port_no
1 to 10 are available for :ROUT:BIAS:PORT.
1, 3, 5, 7 or 9 are available for :ROUT:COUP:PORT.