Diagnostic:test:frame[:execute – Agilent Technologies Agilent E5250A User Manual
Page 186

7-18
Agilent E5250A User’s Guide, Edition 11
Command Reference
:DIAGnostic:TEST:FRAMe[:EXECute]?
:DIAGnostic:TEST:FRAMe[:EXECute]?
This command executes the specified E5250A test, then returns the test result. For
more info, see Chapter 3.
After :DIAG:TEST:FRAM:EXEC CONT execution, the E5250A status becomes
same as after *RST command execution.
After :DIAG:TEST:FRAM:EXEC FPAN execution, the E5250A status is not
changed.
NOTE
The :DIAG:TEST:FRAM:EXEC FPAN command requires key input from the
front panel. If key is not pressed within 10 seconds, this test result will be fail.
Syntax
:DIAGnostic:TEST:FRAMe[:EXECute]? CONTroller | FPANel
Query response
test_result
1 : Fail
0 : Pass
Example
OUTPUT @Hp5250;":DIAG:TEST:FRAM? CONT"
Parameter Explanation
CONTroller
Controller test
FPANel