Agilent E5250A User’s Guide, Edition 11

7-45

Command Reference

:SYSTem:CPON

:SYSTem:CPON

This command resets the specified card to the power-on state. If the parameter is
ALL, all addressable cards are reset to the power-on state. Status after this command
is shown below:

Channel Configuration: No Change

Connection Rule:

Free

Connection Sequence:

Break Before Make

Bias Mode:

Off

Bias Port:

E5252A : 10

E5255A : None

Bias Enabled Channels: All channels (output ports) are bias enabled

Couple Port Mode:

Off

Couple Port:

No Couple Ports

Card Channel Status:

All relays open in present channel configuration mode

Self-Test Result:

No Change

For the device reset state, see *RST.

Syntax

:SYSTem:CPON card_number

Example

OUTPUT @Hp5250;":SYST:CPON ALL"

Parameter Explanation

card_number

card number:

For Auto Config mode: 0 or ALL

For Normal Config mode: 1, 2, 3, 4, or ALL