Agilent Technologies Agilent E5250A User Manual
Page 59

Agilent E5250A User’s Guide, Edition 11
3-9
Executing Self-Test and Leak Test
Executing Self-Test
430 ENTER
@Hp5250;A
440
IF A=0 THEN GOTO 470
450
DISP "Card 3 failed. Press Continue to test Card 4"
460 GOTO
480
470
DISP "Card 3 passed. Press Continue to test Card 4"
480 PAUSE
490 OUTPUT
@Hp5250;":DIAG:TEST:CARD:EXEC?
4"
500 ENTER
@Hp5250;A
510
IF A=0 THEN GOTO 540
520
DISP "Card 4 failed."
530 GOTO
550
540
DISP "Card 4 passed."
550 END
Line Number
Description
30 to 90
Clears all previous test results.
100 to 170
Executes the Controller Test, then returns test result.
180 to 260
Executes the Front Panel Interface Test, then returns the test
result.
270 to 340
Executes the Relay Test for the card installed in slot 1 of the
E5250A, then returns test result.
350 to 410
Executes the Relay Test for the card installed in slot 2 of the
E5250A, then returns test result.
420 to 480
Executes the Relay Test for the card installed in slot 3 of the
E5250A, then returns test result.
490 to 550
Executes the Relay Test for the card installed in slot 4 of the
E5250A, then returns test result.