beautypg.com

Agilent Technologies Agilent E5250A User Manual

Page 128

background image

5-34

Agilent E5250A User’s Guide, Edition 11

Controlling the E5250A
Using Virtual Front Panel Utility

Figure 5-11

Example of SETUP MEMORY MENU

Agilent E5250A Virtual Front Panel Utility

** SETUP MEMORY MENU **

Current Setup Memory : 1

Setup: Comment

Setup: Comment

1 : Open All

13 : Open All

2 : MOSFET 1

14 : Open All

3 : MOSFET 2

15 : Open All

4 : MOSFET 3

16 : Open All

5 : MOS CAP 1

17 : Open All

6 : MOS CAP 2

18 : Open All

7 : MOS CAP 3

19 : Open All

8 : Open All

20 : Open All

9 : Open All

21 : Open All

10 : Open All

22 : Open All

11 : Open All

23 : Open All

12 : Open All

24 : Open All

Select desired softkey.

To change the E5250A setup to specified setup memory

Select the GOTO SETUP softkey. The following message is displayed:

Input setup memory number to be connected?

Type the desired setup memory number, then press the

Enter

key. Current Setup

Memory changes to the specified setup memory, and the E5250A changes to the new
settings.

If you press

Enter

key without typing a number, the operation is canceled.

To copy a setup memory

Select the COPY SETUP softkey. The following message is displayed:

Input copy src & dest setup memory number: src,dest

Type the desired setup memory number, then press the

Enter

key. This operation

overwrites the setup memory data. For example, to copy setup memory No.1 to
No.5, type as follows:

1,5

If you press

Enter

key without typing numbers, operation is canceled.