E5250a card/mode/port related errors – Agilent Technologies Agilent E5250A User Manual
Page 316

11-12
Agilent E5250A User’s Guide, Edition 11
Error Messages
E5250A Specific Error Messages
E5250A Card/Mode/Port Related Errors
Error
Number
Error Message and Description
3000
Card0 initialization fail
The E5250A may be defective. Contact your nearest Agilent
Technologies service center.
3001
Card1 initialization fail
The E5250A or card installed in slot 1 of the E5250A may be
defective. Or if card is the E5255A, the internal cable connection or
the DIP switch setting may be wrong.
3002
Card2 initialization fail
The E5250A or card installed in slot 2 of the E5250A may be
defective. Or if card is the E5255A, the internal cable connection or
the DIP switch setting may be wrong.
3003
Card3 initialization fail
The E5250A or card installed in slot 3 of the E5250A may be
defective. Or if card is the E5255A, the internal cable connection or
the DIP switch setting may be wrong.
3004
Card4 initialization fail
The E5250A or card installed in slot 4 of the E5250A may be
defective. Or if card is the E5255A, the internal cable connection or
the DIP switch setting may be wrong.
3011
Bad couple port number
For the Input Port No. for Couple Port, only 1, 3, 5, 7, or 9 are
allowed.
3012
Bad bias port number
For Input Port No. for Bias Port, only 1 to 10 are allowed.