2 power supply, 3 backup battery, 3 sensors – Campbell Scientific CR800 and CR850 Measurement and Control Systems User Manual
Page 36: 1 analog sensors

Section 4. Quickstart Tutorial
36
4.1.2.2 Power Supply
The CR800 is powered by a nominal 12 Vdc source. Acceptable power range is 
9.6 to 16 Vdc. 
External power connects through the green POWER IN on the face of the 
CR800. The POWER IN connection is internally reverse-polarity protected. 
4.1.2.3 Backup Battery
A lithium battery backs up the CR800 clock, program, and memory in case of 
power loss. See Internal Battery
(p. 76).
4.1.3 Sensors
Most electronic sensors, whether or not manufactured or sold by Campbell 
Scientific, can be interfaced to the CR800. Check for on-line content concerning 
interfacing sensors at www.campbellsci.com, or contact a Campbell Scientific 
applications engineer for assistance. 
4.1.3.1 Analog Sensors
Analog sensors output continuous voltages that vary with the phenomena 
measured. Analog sensors connect to analog terminals. Analog terminals are 
configured as single-ended, wherein sensor outputs are measured with respect to 
ground (figure Analog Sensor Wired to Single-ended Channel #1
(p. 36)
) or
configured as differential, wherein the high output of a sensor is measured with 
respect to the low output (figure Analog Sensor Wired to Differential Channel #1
). Table Single-ended and Differential Input Channels
(p. 37)
lists channel
assignments.
Figure 3: Analog sensor wired to single-ended channel #1
