Section 7. Installation

165

 

'//////////////////////////// PROGRAM ////////////////////////////

BeginProg

'Set Gage Factors

GF_Raw = 2.1
GF_Adj = GF_Raw

'The adj Gage factors are used in the calculation of uStrain

'If a calibration has been done, the following will load the zero or

'Adjusted GF from the Calibration file

LoadFieldCal

(True)

Scan

(100,mSec,100,0)

'Measure Bridge Resistance

BrFull

(Raw_mVperV,1,mV25,1,Vx1,1,2500,True ,True ,0,250,1.0,0)

'Calculate Strain for 1/4 Bridge (1 Active Element)

StrainCalc

(microStrain,1,Raw_mVperV,Zero_mVperV,1,GF_Adj,0)

'Steps (1) & (3): Zero Calibration

'Balance bridge and set Zero_Mode = 1 in numeric monitor. Repeat after

'shunt calibration.

FieldCalStrain

(10,Raw_mVperV,1,0,Zero_mVperV,Zero_Mode,0,1,10,0 ,microStrain)

'Step (2) Shunt Calibration

'After zero calibration, and with bridge balanced (zeroed), set

'KnownRes = to gage resistance (resistance of gage at rest), then set

'Shunt_Mode = 1. When Shunt_Mode increments to 3, position shunt resistor

'and set KnownRes = shunt resistance, then set Shunt_Mode = 4.

FieldCalStrain

(13,MicroStrain,1,GF_Adj,0,Shunt_Mode,KnownRes,1,10,GF_Raw,0)

CallTable

CalHist

NextScan

EndProg

 

7.8.1.6.1 Quarter-Bridge Shunt (Option 13)

With CRBasic example FieldCalStrain() Calibration Demo

(p. 164)

sent to the

CR800, and the strain gage stable, use the external keyboard / display or software
numeric monitor to change the value in variable KnownRes to the nominal
resistance of the gage, 1000 Ω, as shown in figure Strain-Gage Shunt Calibration
Started

(p. 166).

Set Shunt_Mode to 1 to start the two-point shunt calibration.

When Shunt_Mode increments to 3, the first step is complete.

To complete the calibration, shunt R1 with the 249-kΩ resistor. Set variable
KnownRes to 249000. As shown in figure Strain-Gage Shunt Calibration
Finished

(p. 166),

set Shunt_Mode to 4. When Shunt_Mode = 6, shunt calibration

is complete.