Campbell Scientific CR800 and CR850 Measurement and Control Systems User Manual
Page 155

Section 7. Installation
155
1. Send CRBasic example FieldCal Zeroing Demonstration Program
(p. 155)
to the
CR800. An excitation channel has been programmed to simulate a sensor
output.
2. To place the simulated RH sensor in a simulated-calibration condition (in the
field it would be placed in a desiccated chamber), place a jumper wire between
channels VX1/EX1 and SE6 (3L). Set variable mV to 1000. Set variable
KnownRH to 0.0.
3. To simulate a calibration, change the value in variable CalMode to 1 to start
calibration. When CalMode increments to 6, zero calibration is complete.
Calibrated Offset will equal -50% at this stage of this example.
Figure 48: Zero (Option 0)
4. To continue this example and simulate a zero-drift condition, change variable
mV to 1050.
5. To simulate conditions for a 30-day, service-calibration, again with desiccated
chamber conditions, set variable KnownRH to 0.0. Change the value in
variable CalMode to 1 to start calibration. When CalMode increments to 6,
simulated 30-day, service zero calibration is complete. Calibrated Offset will
equal -52.5%.
CRBasic Example 26. FieldCal() Zeroing Demonstration Program
'Jumper VX1/EX1 to SE6(3L) to simulate a sensor
Public
mV
'Excitation mV Output
Public
KnownRH
'Known Relative Humidity
Public
CalMode
'Calibration Trigger
Public
Multiplier
'Multiplier (Starts at .05 mg / liter / mV,
'does not change)
Public
Offset
'Offset (Starts at zero, not changed)
Public
RH
'Measured Relative Humidity
'Data Storage Output of Calibration Data -- stored whenever a calibration occurs
DataTable
(CalHist,NewFieldCal,200)
SampleFieldCal
EndTable
BeginProg
Multiplier = .05
Offset = 0
LoadFieldCal
(true)
'Load the CAL File, if possible