Section 8. Operation

280 

 

 

duration. Consequently, noise at 1 / (integer multiples) of the integration duration
is effectively rejected by an analog integrator. table CRBasic Measurement
Integration Times and Codes

(p. 280)

lists three integration durations available in the

CR800 and associated CRBasic codes. If reversing the differential inputs or
reversing the excitation is specified, there are two separate integrations per
measurement; if both reversals are specified, there are four separate integrations.

Table 54. CRBasic Measurement Integration Times and Codes

Integration Time (ms)

CRBasic Code

Comments

250 µs

250

Fast integration

16.667 ms

_60Hz

Filters 60-Hz noise

20 ms

_50Hz

Filters 50-Hz noise

 

8.1.2.7.1 ac Power Line Noise Rejection

Grid or mains power (50 or 60 Hz, 230 or 120 Vac) can induce electrical noise at
integer multiples of 50 or 60 Hz. Small analog voltage signals, such as
thermocouples and pyranometers, are particularly susceptible. CR800 voltage
measurements can be programmed to reject (filter) 50-Hz or 60-Hz related noise.

ac Noise Rejection on Small Signals

The CR800 rejects ac power line noise on all voltage ranges except

mV5000

and

mV2500

by integrating the measurement over exactly one ac cycle before A/D

conversion as illustrated in table ac Noise Rejection on Small Signals

(p. 280)

and

the full cycle technique of figure ac Power Line Noise Rejection Techniques

(p.

281).

Table 55. ac Noise Rejection on Small Signals

Applies to all analog input voltage ranges except

mV2500

and

mV5000

.

ac Power Line Frequency

Measurement Integration

Duration

CRBasic Integration Code

60 Hz

16.667 ms

_60Hz

50 Hz

20 ms

_50Hz