Section 7. Installation

166 

 

 

Figure 50: Strain-gage shunt calibration started

Figure 51: Strain-gage shunt calibration finished

7.8.1.6.2 Quarter-Bridge Zero (Option 10)

Continuing from Quarter-Bridge Shunt (Option 13)

(p. 165),

keep the 249-kΩ

resistor in place to simulate a strain. Using the external keyboard / display or
software numeric monitor, change the value in variable Zero_Mode to 1 to start
the zero calibration as shown in figure Starting Zero Procedure (p. 166). When
Zero_Mode increments to 6, zero calibration is complete as shown in figure Zero
Procedure Finished

(p. 167).

Figure 52: Starting zero procedure