9 field calibration, 10 cabling effects, 1 analog sensor cables – Campbell Scientific CR800 and CR850 Measurement and Control Systems User Manual
Page 309: 2 pulse sensors, 309) sh

Section 8. Operation
309
Figure 98: Circuit to limit control port input to 5 Vdc
8.1.9 Field Calibration
Read More! Field Calibration of Linear Sensors (FieldCal)
(p. 151)
has complete
information.
Calibration increases accuracy of a measurement device by adjusting its output, or 
the measurement of its output, to match independently verified quantities. 
Adjusting a sensor output directly is preferred, but not always possible or 
practical. By adding FieldCal() or FieldCalStrain() instructions to the CR800 
program, a user can easily adjust the measured output of a linear sensors by 
modifying multipliers and offsets. 
8.1.10 Cabling Effects
Sensor cabling can have significant effects on sensor response and accuracy. This 
is usually only a concern with sensors acquired from manufacturers other than 
Campbell Scientific. Campbell Scientific sensors are engineered for optimal 
performance with factory-installed cables. 
8.1.10.1 Analog Sensor Cables
Cable length in analog sensors is most likely to affect the signal settling time. For 
more information, see Signal Settling Time
8.1.10.2 Pulse Sensors
Because of the long interval between switch closures in tipping bucket rain gages, 
appreciable capacitance can build up between wires in long cables. A built-up 
charge can cause arcing when the switch closes, shortening switch life. As shown 
in figure Current Limiting Resistor in a Rain Gage Circuit
a 100-ohm
resistor is connected in series at the switch to prevent arcing. This resistor is 
installed on all rain gages currently sold by Campbell Scientific. 
Figure 99: Current limiting resistor in a rain gage circuit
