Figure 48: zero (option 0) – Campbell Scientific CR800 and CR850 Measurement and Control Systems User Manual

Page 158

background image

Section 7. Installation

158 

 

 

K =  temperature correction coefficient (‐0.04 PSI / C° is typical) 

T

0

 = r temperature at the zero state 

T

1

 =  temperature measurement 

S

0

 = barometric pressure at the zero state 

S

1

 = barometric pressure measurement. 

 

The following procedure determines zero offset of the pressure transducer, water
temperature, and barometric pressure readings. Use the external keyboard /
display or support software numeric monitor to change variable values as directed.

Calibration Report for Pressure Transducer


Parameter

Measurement
Before Zero

Measurement
After Zero

Piezometer Output (digits)

8746

0

Piezometer Temperature (°C)

21.4

0

Barometer Pressure (mb)

991

0

 

1. Send CRBasic example FieldCal() Zero Basis Demo Program

(p. 158)

to the

CR800.

2. To simulate the pressure transducer in zero conditions:

Digits_Measured is set to 8746 automatically

Temp_Measured is set to 21.4 automatically

BP_Measured to 991

3. To simulate the calibration, change the value in variable CalMode to 1 to start

calibration. When CalMode increments to 6, zero calibrations are complete.
Calibrated offsets will equal 8746 digits, 21.5 C°, and 991 mb.

CRBasic Example 28. FieldCal() Zero Basis Demo Program 

'FieldCal zero basis demonstration program

Public

Pressure

Public

VW(1,6)

Public

Equation_Parameters(3)

Alias

Equation_Parameters(1) = Digits_Measured

Alias

Equation_Parameters(2) = Temp_Measured

Alias

Equation_Parameters(3) = BP_Measured