beautypg.com

2 two-point calibrations (multiplier / gain) – Campbell Scientific CR800 and CR850 Measurement and Control Systems User Manual

Page 153

background image

Section 7. Installation

153

 

topics. The most comprehensive resource to date covering use of FieldCal() and
FieldCalStrain() is RTDAQ software documentation. Be aware that,

• the CR800 does not check for out-of-bounds values in mode variables.
• valid mode variable entries are "1" or "4".

7.8.1.4.1 Single-Point Calibrations (zero, offset, or zero basis)

Use this single-point calibration procedure to adjust an offset (y-intercept). See
Zero (Option 0)

(p. 156)

and Offset (Option 1)

(p. 156)

for demonstration programs:

1. Use a separate FieldCal() instruction and separate field cal variables for each

sensor to be calibrated. In the CRBasic program , put the FieldCal()
instruction immediately below the associated measurement instruction.

2. Set mode variable = 0 or 6 before starting.

3. Place the sensor into zeroing or offset condition.

4. Set

KnownVar

variable to the offset or zero value.

5. Set mode variable = 1 to start calibration.

Mode Variable

Interpretation

> 0 and ≠ 6

calibration in progress

< 0

calibration encountered an error

2

calibration in process

6

calibration complete.

 

7.8.1.4.2 Two-point Calibrations (multiplier / gain)

Use this two-point calibration procedure to adjust multipliers (slopes) and offsets
(y-intercepts). See Two Point Slope and Offset (Option 2)

(p. 159)

and Two Point

Slope Only (Option 3)

(p. 161)

for demonstration programs:

1. Use a separate FieldCal() instruction and separate, related variables for each

sensor to be calibrated.

2. Ensure mode variable = 0 or 6 before starting.

a.

If

Mode > 0 and ≠ 6, calibration is in progress.

b.

If

Mode < 0, calibration encountered an error.

3. Place sensor into first known point condition.

4. Set

KnownVar

variable to first known point.

5. Set

Mode

variable = 1 to start first part of calibration.

a.

Mode

= 2 (automatic) during the first point calibration.

b.

Mode

= 3 (automatic) when the first point is completed.

6. Place sensor into second known point condition.

7. Set

KnownVar

variable to second known point.