2 two-point calibrations (multiplier / gain) – Campbell Scientific CR800 and CR850 Measurement and Control Systems User Manual
Page 153

Section 7. Installation
153
topics. The most comprehensive resource to date covering use of FieldCal() and
FieldCalStrain() is RTDAQ software documentation. Be aware that,
• the CR800 does not check for out-of-bounds values in mode variables.
• valid mode variable entries are "1" or "4".
7.8.1.4.1 Single-Point Calibrations (zero, offset, or zero basis)
Use this single-point calibration procedure to adjust an offset (y-intercept). See
Zero (Option 0)
(p. 156)
and Offset (Option 1)
(p. 156)
for demonstration programs:
1. Use a separate FieldCal() instruction and separate field cal variables for each
sensor to be calibrated. In the CRBasic program , put the FieldCal()
instruction immediately below the associated measurement instruction.
2. Set mode variable = 0 or 6 before starting.
3. Place the sensor into zeroing or offset condition.
4. Set
KnownVar
variable to the offset or zero value.
5. Set mode variable = 1 to start calibration.
Mode Variable
Interpretation
> 0 and ≠ 6
calibration in progress
< 0
calibration encountered an error
2
calibration in process
6
calibration complete.
7.8.1.4.2 Two-point Calibrations (multiplier / gain)
Use this two-point calibration procedure to adjust multipliers (slopes) and offsets
(y-intercepts). See Two Point Slope and Offset (Option 2)
(p. 159)
and Two Point
Slope Only (Option 3)
(p. 161)
for demonstration programs:
1. Use a separate FieldCal() instruction and separate, related variables for each
sensor to be calibrated.
2. Ensure mode variable = 0 or 6 before starting.
a.
If
Mode > 0 and ≠ 6, calibration is in progress.
b.
If
Mode < 0, calibration encountered an error.
3. Place sensor into first known point condition.
4. Set
KnownVar
variable to first known point.
5. Set
Mode
variable = 1 to start first part of calibration.
a.
Mode
= 2 (automatic) during the first point calibration.
b.
Mode
= 3 (automatic) when the first point is completed.
6. Place sensor into second known point condition.
7. Set
KnownVar
variable to second known point.