beautypg.com

Figure 80: pt100 in three-wire half-bridge – Campbell Scientific CR800 and CR850 Measurement and Control Systems User Manual

Page 266

background image

Section 7. Installation

266 

 

 

The recorded amplitude for this example should be about 1/3 of the 
input‐signal amplitude.  A program was written with two stored 
variables: Accel2 and Accel2RA.  The raw measurement was stored in 
Accel2, while Accel2RA was the result of performing a running average 
on the Accel2 variable. Both values were stored at a rate of 500 Hz. 
Figure Running‐Average Signal Attenuation

 (p. 267) 

show the two values 

plotted in a single graph to illustrate the attenuation (the running‐
average value has the lower amplitude). 

The resultant delay (delay in time) = (Scan rate)(N‐1)/2 = 2 ms (10‐1)/2 = 
9 ms.  This is about 1/3 of the input‐signal period. 

Figure 83: Running-average frequency response