beautypg.com

Readback commands – GW Instek PPH-1503 User Manual

Page 95

background image

REMOTE CONTROL

95

Readback Commands

Command

:SENSe[1]:FUNCtion

Function

Selects the type of measurement function: voltage,
current, pulse, long integration and DVM
measurement.

Description

name

“VOLTage”: Voltage measurement.
“CURRent”: Current measurement.
“PCURrent”: Pulse current
measurement.
“LINTegration”: Long integration
measurement.
“DVMeter”: DVM input measurement.

Example

:SENSe:FUNCtion “VOLTage”
Selects “Voltage” as the measurement type.

Command

:SENSe[1]:FUNCtion?

Function

Queries the type of measurement function.

Response time Maximum: 16ms
Example

:SENSe:FUNCtion?
Returns the type of measurement function.


Command

:SENSe[1]:NPLCycles

Function

Sets the number of PLCs for the integration rate
for voltage, current and DVM measurements.

Description

0.01-10.00

Example

:SENSe:NPLCycles 0.10
Sets the number of PLCs to 0.1.