TrigLeve[3]

Trigger Level.

To avoid false pulse
measurements, the trigger level
can be set close to the current
amplitude. All noise and transient
currents that are below the trigger
level will be ignored. The trigger
level has a setting range of 0 to 5A,
with a resolution of 5mA. The
setting unit for the trigger level is
in amps (A). This setting is only
valid for pulse measurements.

RecallSetup

Recalls stored settings. A total of 6
settngs can be recalled:

RST/SAV0 to SAV4

Parameter
Settings

IntTime

Use the arrow keys to select

IntTime

,

press Enter and then use the arrow
keys to select the type of integration
time that you want to set (Hight
Time, Low Time, Aver Time). Press
Enter again to set the integration
time. After inputing the integration
time, press Enter to return to the
pulse current measurement menu. If

Auto Time

was selected, press Enter to

return to the pulse current
measurement menu.

Example:

High Time 33uS:

IntTime



Hight Time



using the