REMOTE CONTROL
129
Errors
Error Message
•Errors are stored in a first in-first out (FIFO) order. The first error message
that is returned is the first error message that was stored. When an error is
read it is also cleared from the queue.
•If there are more than 10 errors produced the last error in the queue is
replaced with “Que overflow”. Unless the error queue is cleared, no more
errors can be written to the error queue. If there are no errors in the error
queue, the instrument will return “No error”.
•To clear the error queue, you can use the
:SYSTem:CLEar command or
cycle the power. When you read a message from the error queue that
message will be cleared from the error queue. Using the *RST command
to reset the instrument does not clear the error queue.
•Remote control instructions can be used to clear the error queue. See the
instructions listed in the previous chapter for details.
Command Errors
-440 Query unterminated after indefinite
-430 Response
-420 Query deadlocked
-410 Query unterminated
-363 Query interrupted
-350 Input buffer overrun
-330 Queue overflow
-314 Self-test failed
-315 Save/recall memory lost
-260 Configuration memory lost
-241 Expression error