beautypg.com

GW Instek PPH-1503 User Manual

Page 102

background image

PPH-1503 User Manual

102

Command

:SENSe[1]:PCURrent:SYNChronize:TLEVel?

Function

Queries the trigger level.

Example

:SENSe:PCURrent:SYNChronize:TLEVel?
Returns the trigger level.


Command

:SENSe[1]:LINTegration:TIME

Function

Sets the long integration integration time.

Description

X (power line frequency: X=0.840 for
50Hz, X=0.850 for 60Hz)

Example

:SENSe:LINTegration:TIME 1.2
Sets the long integration time to 1.2S.


Command

:SENSe[1]:LINTegration:TIME?

Function

Queries the the long integration time.

Example

:SENSe:LINTegration:TIME?
Returns the long integration time.


Command

:SENSe[1]:LINTegration:TIME:AUTO

Function

Sets the long itegration time to the auto setting.

Example

:SENSe:LINTegration:TIME:AUTO
Sets the long itegration time to the auto setting.


Command

:SENSe[1]:LINTegration:TLEVel

Function

Sets the long integration trigger level.

Description

0.000-5.000