beautypg.com

System related commands – GW Instek PPH-1503 User Manual

Page 118

background image

PPH-1503 User Manual

118

System Related Commands

Command

*IDN?

Function

Read the instrument identification .

Description

The return string contains four field,

each separated by a comma. The first
field is the manufacturer, followed by
the model name, serial number and the
version number.

Example

*IDN?
Returns: GW,PPH

-

1503,XXXXXXXX,V0.62

GW: Manufacturer,
PPH

-

1503: Model name,

XXXXXXXX: Serial number,
V0.62: version number.

Command

*RST

Function

Resets the unit to RST default conditions.

Example

*RST
Resets the unit.

Command

*TST?

Function

Performs checksum test on the RAM.

Return
value

0: No errors
2: Indicates that there is a RAM error.

Example

*TST?
Return 0 if there are no errors, returns 2 if there is
an error.