CONFIGURATION

97

Configuration

Configuration of the PSW power supplies is divided into five
different configuration settings: Normal Function, USB/GPIB, LAN,
Power ON Configuration, Calibration Settings and System Settings.
Power ON Configuration differs from the other settings in that the
settings used with Power ON Configuration settings can only be set
during power up. The other configuration settings can be changed
when the unit is already on. This prevents some important
configuration parameters from being changed inadvertently. Power
On Configuration settings are numbered F-90 to F-95 and the other
configuration settings are numbered F-00 to F-61 and F-88 to F-89.

Configuration Table

Please use the configuration settings listed below when applying
the configuration settings.

Normal Function
Settings

Setting Setting Range

Output ON delay time

F-01

0.00s~99.99s

Output OFF delay time F-02

0.00s~99.99s

V-I mode slew rate select F-03

0 = CV high speed priority
1 = CC high speed priority
2 = CV slew rate priority
3 = CC slew rate priority

Rising voltage slew rate F-04

0.01V/s~60.00V/s (PSW 30-XX)
0.1V/s~160.0V/s (PSW 80-XX)
0.1V/s~320.0V/s (PSW 160-XX)
0.1V/s~500.0V/s (PSW 250-XX)
1V/s~1600V/s (PSW 800-XX)

Falling voltage slew rate F-05

0.01V/s~60.00V/s (PSW 30-XX)
0.1V/s~160.0V/s (PSW 80-XX)
0.1V/s~320.0V/s (PSW 160-XX)
0.1V/s~500.0V/s (PSW 250-XX)
1V/s~1600V/s (PSW 800-XX)