Master-Slave Parallel Operation

Master-Slave
Configuration

Before using the power supplies in parallel, the
master and slave units need to be configured.

Steps

1. Configure the OVP and OCP

settings for the master unit.

Page 57

2. For each unit, hold the Function

key while turning the power on to
enter the power on configuration
settings.

A

W

V

W

%W

100

80

60

40

20

ISR

C C

DLY

ALM

RMT

C V

VSR

Function

OVP/OCP

Set

Output

PWR DSPL

Lock/Local

Test

PSW 30-36

360W

Voltage

Current

Multi-Range DC Power Supply

3. Configure F-93 (Master/Slave)

setting for each master/slave unit.

Page 114

Unit

F-93

Master (with 1 slave in parallel)

1

Master (with 2 slaves in parallel)

2

Slave unit (parallel slave)

3

4. Cycle the power on the units (reset the power).

Note

Configuration settings can be checked for both the
master and slave units by pressing the Function
key and checking F-93.

Only the Master OVP and OCP level is used for
over voltage and current protection. Slave OVP and
OCP level is disregarded.

OTP works independently for each unit.