beautypg.com

GW Instek PSW-Series User Manual User Manual

Page 58

background image

PSW Series User Manual

58

ALM

V

A

Before setting the OVP or OCP level:

Ensure the load is not connected.

Ensure the output is set to off.

Setting Ranges

PSW (360W)

30-36

80-13.5 160-7.2 250-4.5 800-1.44

OVP Range (V)

3-33

8-88

16-176

20-275

20-880

OCP Range (A)

3.6-39.6

1.35-14.85 0.72-7.92

0.45-4.95

0.144-1.584

PSW (720W)

30-72

80-27

160-14.4 250-9

800-2.88

OVP Range (V)

3-33

8-88

16-176

20-275

20-880

OCP Range (A)

5-79.2

2.7-29.7

1.44-15.84 0.9-9.9

0.288-3.168

PSW (1080W)

30-108

80-40.5 160-21.6 250-13.5 800-4.32

OVP Range (V)

3-33

8-88

16-176

20-275

20-880

OCP Range (A)

5-118.8

4.05-44.55 2.16-23.76 1.35-14.85 0.432-4.752

Steps

1. Press the OVP/OCP key. The

OVP/OCP key lights up.

OVP/OCP

2. The OVP setting will be displayed on the top

and the OCP setting (or OFF) will be displayed
on the bottom.

V

A

OVP Setting