GW Instek PSW-Series User Manual User Manual
Page 118
PSW Series User Manual
118
D COM
2 Connected to the (–S) sense- terminal when
remote sense is used. Connected to the negative
output terminal when remote sense is not used.
CURRENT SUM
OUT
3 Current sum output signal when used in parallel
mode.
EXT-V CV CONT 4 External voltage control of the voltage output. A
voltage of 0~10V is used to control the full scale
voltage output (0%~100%) of the instrument.
EXT-V CC CONT 5 External voltage control of the current output. A
voltage of 0~10V is used to control the full scale
current output (0%~100%)of the instrument
EXT-R CV CONT
PIN1
6 External resistance control of the voltage output. A
resistance of 0kΩ ~ 10kΩ is used to control the full
scale voltage output (0%~100%) of the instrument.
EXT-R CV CONT
PIN2
7 External resistance control of the voltage output. A
resistance of 0kΩ ~ 10kΩ is used to control the full
scale voltage output (0%~100%) of the instrument.
EXT-R CC CONT
PIN1
8 External resistance control of the current output. A
resistance of 0kΩ ~ 10kΩ is used to control the full
scale current output (0%~100%) of the instrument.
EXT-R CC CONT
PIN2
9 External resistance control of the current output. A
resistance of 0kΩ ~ 10kΩ is used to control the full
scale current output (0%~100%) of the instrument.
V MON
10 Voltage Monitor Output. Outputs the full scale
voltage (0~100%) as a voltage (0V~10V).
I MON
11 Current Monitor Output. Outputs the full scale
current (0~100%) as a voltage (0V~10V).
SHUTDOWN
12 The shut down signal will turn off the output or
power when a low TTL signal is applied. The
shutdown signal is pulled up to 5V with a 10kΩ
pull-up resistor.
CURRENT_SUM_
1
13 Master unit current sum input signal from first
slave CURRENT SUM OUTPUT. Used in parallel
mode only.
CURRENT_SUM_
2
14 Master unit current sum input signal from second
slave CURRENT SUM OUTPUT. Used in parallel
mode only.
FEEDBACK
15 Parallel control signal during master-slave parallel
operation.