4. Press the Output key. The voltage

can now be controlled with the
External resistance.

Output

Note

Ensure the resistor(s) and cables used exceed the
isolation voltage of the power supply. For example:
insulation tubes with a withstand voltage higher
than the power supply can be used.

When choosing an external resistor ensure the
resistor can withstand a high degree of heat.

Note

CV and CC Slew Rate Priority are disabled for V-I
mode (F-03) when using external resistance
control. See the normal function settings on page
102.

External Resistance Control of Current Output

Background

External resistance control of the current output is
accomplished using the MIL-26 connector on the
rear panel. A resistance of 0kΩ~10kΩ is used to
control the full scale current of the instrument.

The output current (0 to full scale) can be
controlled with the external resistance going up
(Ext-R ) 0kΩ~10kΩ(10kΩ = Vo,max) or down
(Ext-R

) 10kΩ~0kΩ(10kΩ = 0).

For 0kΩ~10kΩ: Output current = full scale current
× (external resistance/10)

For 10kΩ~0kΩ: Output current = full scale current
× ([10-external resistance]/10)