Set to c.c. mode – GW Instek PSW-Series User Manual User Manual

Page 63

background image

OPERATION

63

Set to C.C. Mode

When setting the power supply to constant current mode, a voltage
limit must also be set to determine the crossover point. When the
voltage exceeds the crossover point, the mode switches to C.V.
mode. For details about C.C. operation, see page 22. C.C. and C.V.
mode have two selectable slew rates: High Speed Priority and Slew
Rate Priority. High Speed Priority will use the fastest slew rate for
the instrument while Slew Rate Priority will use a user-configured
slew rate.

Background

Before setting the power supply to
C.C. mode, ensure:

The output is off.

The load is connected.

Steps

1. Press the Function key. The

Function key will light up.

Function

2. The display should show F-

01 on the top and the
configuration setting for
F-01 on the bottom.

3. Rotate the Voltage knob to change

the F setting to F-03 (V-I Mode
Slew Rate Select).

Voltage

4. Use the Current knob to set the F-

03 setting.

Set F-03 to 1 (CC High Speed
Priority) or 3 (CC Slew Rate
Priority) and save.

Current