7 resetting historical data, Resetting historical data -18 – Yokogawa GC1000 Mark II Process Gas Chromatograph User Manual
Page 93
<6. Analysis Results Windows>
6-18
IM 11B03G03-03E
6.7
Resetting Historical Data
If you want to record data from scratch in the Concentration Analysis History and Retention Time
History windows, execute the [Reset History] command. Executing the command simultaneously
erases all data in both windows, regardless of the state of the analysis results windows.
n
Resetting Historical Data in the Concentration Analysis History and
Retention Time History Windows
l
Procedure
(1) Open the Latest Analysis Results window and click on the [Reset History] command in the
[File] menu. A dialog box pops up to ask you if you want to reset the historical data.
F0618.ai
F0619.ai
(2) Click on the [OK] button. This erases all data in the Concentration Analysis History and
Retention Time History windows.
•
The heading “Strm#-Peak#-Peak Name” remains in the windows.
NOTE
If you have made any change to the operation pattern or peak settings while the Analysis Results
window is active:
• Execute the [Exit] command in the [File] menu to close the Analysis Results window once.
Next, exit the Maintenance Terminal. And then restart the Maintenance Terminal, and
reopen the Analysis Results window. Executing the [Reset History] command without exiting
the Analysis Results window results in incorrect data display.