<3. Analyzer Operation Window>
3-21
IM 11B03G03-03E
n
Uploading
To upload and save parameter settings from the analyzer to a disk of the Maintenance Terminal,
take the following procedure.
l
Procedure
(1) Click on the [Upload to PC] command in the [File] menu. The Save As dialog box pops up
where you can specify the file name for the parameter setting information to be uploaded.
The default directory is “Data” right under the installation directory.
Input
F0336.ai
(2) Type the name of the file where you want to save the parameter settings, with an extension
of “.pr6” in the “File name” field.
Example: When the file name is “system01”, type
system01.pr6
(3) Click on the [OK] button. A dialog box pops up to show the status of the transmission of the
parameter settings from the analyzer.
F0337.ai
When all parameter settings have been uploaded and saved in the file, the following dialog
box pop up.
F0338.ai
• It takes about 40 seconds for data transmission from the analyzer (if the communication
speed is 38400 bps).
(4) Click on the [OK] button. The uploading has been completed.
HELP !
To interrupt the uploading of parameter settings;
• Click on the [Interrupt] button in the dialog box showing the transmission status. A confirmation dialog box
pops up.
• Click on the [Yes] button.