Ntypes and change of the measurement status, Ltypes of the measurement status, Lchanging the measurement status – Yokogawa GC1000 Mark II Process Gas Chromatograph User Manual
Page 43: Lcalibration (validation) method

<3. Analyzer Operation Window>
3-14
IM 11B03G03-03E
n
Types and Change of the Measurement Status
l
Types of the Measurement Status
There are four types of the measurement status, as shown below.
Measurement
Button
Description
Stream Sequence
Performs stream sequence measurement.
Stream (Continueous)
Performs continuous measurements of the selected
stream.
Stream (1 cycle)
Performs a single measurement of the selected stream.
Calibration (Validation)
Performs calibration or validation.
The current state of the measurement status is indicated in the bottom half of the Operation
Status ribbon and each button is only used for executing commands.
l
Changing the Measurement Status
To change the measurement status, click on the button of the measurement status you want to
select. Depending on the current situation, operable buttons, which are indicated as darkened
buttons, are different.
There is a [Cancel] command button
which is only used to cancel the next measurement.
l
Calibration (Validation) Method
There are three types of the calibration (validation) method, as shown below.
Calibration
(Validation) Method
Description
Automatic
Performs calibration (validation) automatically according
to the starting time and intervals specified in the analyzer.
Semi-automatic
Performs calibration (validation) by clicking on the Cal
(Val) button and specifying the number.
Manual
Performs calibration validation) manually by clicking on
the Cal (Val) button, arranging the conditions, and then
clicking on the start button
and end button
.
Take the following procedure to select the calibration (validation) method.
(1) Click on the [Calibration/Validation Method] button
. The Calibration/Validation Method
dialog pops up.
F0319.ai
4th Edition : Mar. 21, 2012-00