beautypg.com

Filter time constant, Sensor selection, Set point low limit and high limit – Watlow EZ-ZONE PM User Manual

Page 78

background image

Watlow EZ-ZONE

®

PM Limit Controller

75

Chapter 8 Features

10. Read Electrical Measurement value [`Mu] of controller via EZ-Configurator or RUI. This will be re-

ferred to as Electrical Measured High. Record high value ______________

11. Calculated Electrical Input Slope = (Precision High – Precision Low) / (Electrical Measured High - Elec-

trical Measured Low). Calculated Slope value ___________

12. Calculated Electrical Input Offset = Precision Low – (Electrical Input Slope * Measured Low). Calculated

Offset value ___________

13. Enter the calculated Electrical Input Slope [ELi;S] and Electrical Input Offset [ELi;o] into the control-

ler.

14. Exit calibration menu.
15. Validate calibration process by utilizing a calibrator to the analog input.
16. Enter calibration offset as recorded in step 2 if required to compensate for sensor error.
Setting Electrical Input Slope [ELi;S] to 1.000 and Electrical Input Offset [ELi;o] to 0.000, restores factory

calibration as shipped from factory.

Filter Time Constant

Filtering smoothes an input signal by applying a first-order filter time constant to the signal. Filtering the

displayed value makes it easier to monitor. Filtering the signal may improve the performance of PID control

in a noisy or very dynamic system.

Adjust the filter time interval with Filter Time [`FiL] (Setup Page, Analog Input Menu).
Example: With a filter value of 0.5 seconds, if the process input value instantly changes from 0 to 100 and re-

mained at 100, the display will indicate 100 after five time constants of the filter value or 2.5 seconds.

Sensor Selection

You need to configure the controller to match the input device, which is normally a thermocouple, RTD or

process transmitter. When you select an input device, the controller automatically sets the input linearization

to match the sensor. It also sets high and low limits, which in turn limit the set point range-high and range-

low values.

Select the sensor type with Sensor Type [`Sen] (Setup Page, Analog Input Menu).

Note:

The EZ-ZONE PM does not have an open-sensor detection feature for process inputs.

Set Point Low Limit and High Limit

The controller constrains the Limit set point to a value between the Set Point Low Limit and the Set Point

High Limit.

Set the set point range with Set Point Low Limit [sp;ll] and Set Point HighLimit [sp;lh] (Setup Page,

Loop Menu).

Unfiltered Input Signal

Time

Temperature

Filter Time Constant

Filtered Input Signal

Time

Temperature