Chapter three – Teledyne LeCroy WaveExpert series Automation Manual User Manual
Page 83
Chapter Three
Control Reference
Sets/Queries the probe attenuation. The probe attenuation is the factor by which the signal is made
smaller, for example, 10 means that the probe divides by 10, and is referred to as a ÷10 probe. Note
that certain passive probes may be marked as 'x10', even though they actually divide the input signal
by a factor of 10.
ProbeAttenuation
From 1e-006 to 10000 step 1e-006
Double
' Visual Basic Script
Set app = CreateObject("LeCroy.XStreamDSO")
' Set the probe attenuation for channel C1 to 100
app.Acquisition.C1.ProbeAttenuation = 100
Range
Example
Description
Queries the name of connected probe.
ProbeName
Any number of characters
String
Range
Description
Sets/Queries the vertical scale (in Volts/Division) of an input channel. When variable gain
(VerScaleVariable control) is disabled, the control will clip values to a 1..2..5 sequence. When variable
gain is enabled, the setting resolution lies in the range 1% to 2%, depending upon the numerical value.
VerScale
From 0.001 to 1 step 0.0005, locked to 1 2 5, fine grain allowed=true, on=false
DoubleLockstep
' Visual Basic Script
Set app = CreateObject("LeCroy.XStreamDSO")
' Set C1 to a scale of 250mV/Div in Variable Scale mode
app.Acquisition.C1.VerScaleVariable = True
app.Acquisition.C1.VerScale = 0.25
Range
Example
Description
3-16
916435 RevA