beautypg.com

Chapter four – Teledyne LeCroy WaveExpert series Automation Manual User Manual

Page 253

background image

Chapter Four

Control Reference

When this control is set to true (default is false), the TIE values the result at the output have "virtual
edges" included in the output (i.e. values which are linearly interpolated) corresponding to edges which
did NOT transit. This allows a relatively uniform in time strm of values. This feature is mostly obviated
by the processor "ParamUpSample"

IncludeVirtualEdges

Bool

Description

This should be set to "true" for TIE analysis of an NRZ data stream. It should be set to false for TIE
analysis of a clock signal

DataIsNRZ

Bool

Description

Timing Analysis can either performed using the edge-edge timing methodology (as was developed for
Time-Interval-Analyzers", or edge-ref, as is common for real-time oscilloscopes. Edge-Ref is highly
recommended.

IntervalType

Enum

Description

Values

EDGEEDGE

EDGEREF

For edge-edge methodology, this control sets the number of UI ( unit intervals) between edges to be
analyzed.

IntervalsEdgeEdge

From 1 to 500 step 1

Integer

Range

Description

Summary

Any number of characters

String

Range

4-49