Chapter three, Control reference – Teledyne LeCroy WaveExpert series Automation Manual User Manual
Page 101
Chapter Three
Control Reference
countermanded. "Auto" allows the instrument to set the grid mode most suitable for the current
number of visible traces.
' Visual Basic Script
Set app = CreateObject("LeCroy.XStreamDSO")
' Enter Octal grid mode
app.Display.GridMode = "Octal"
Example
Values
Auto
Automatically choose grid mode, one trace per grid
Dual
Dual grid mode
Octal
Octal grid mode
Quad
Quad grid mode
Single
Single grid mode
XY
XY grid mode
XYDual
XY + Dual grid mode
XYSingle
XY + Single grid mode
Sets/Queries the style in which traces are drawn.
TraceStyle
Enum
' Visual Basic Script
Set app = CreateObject("LeCroy.XStreamDSO")
' Read the state of the persistence mode.
TraceStyle = app.Display.TraceStyle
Example
Description
Values
Line
Connect adjacent samples with straight lines
Points
Show only the sample points
Sets/Queries the grid intensity as a percentage of the maximum value, with a resolution of 1%.
GridIntensity
From 0 to 100 step 1
Integer
' Visual Basic Script
Set app = CreateObject("LeCroy.XStreamDSO")
' Set the grid intensity to 60% of the maximum.
app.Display.GridIntensity = 60
Range
Example
Description
3-34
916435 RevA