beautypg.com

Status – Teledyne LeCroy WaveExpert series Automation Manual User Manual

Page 58

background image

A

BOUT

R

ESULTS

Status

Description: Result status. Status is returned as a decimal value, but should be treated like a register with the

following bit values. See the example for decoding the Status result.

64-bit hexadecimal

int value shown in

XStreamBrowser

VT_DECIMAL

value

returned

from Status

property

StatusDescription

Description

0x1 0.0001 Invalid

result

Indicate that this Result cannot

be used for any calculation

0x2 0.0002

Data

overflow

Result contains Overflow data or

is calculated from a Result

containing Overflows.

0x4 0.0004

Data

underflow

Result contains Underflow data

or is calculated from a Result

containing Underflows

0x8 0.0008

Some values are

undefined

Result contains some values or

data that are undefined.

0x10 0.0016 Less

than

Actual value is likely to be less

than the calculated value. Mostly

used for parameter

measurements.

0x20 0.0032 Greater

than

Actual value is likely to be

greater than the calculated value.

Mostly used for parameter

measurements.

0x40

0.0064

Not a pulse

Analyzed signal is not recognized

as a pulse; mostly used for pulse

parameter measurements.

0x80 0.0128 Not

cyclic

Contains some values measured

on non-cyclic signal.

0x100 0.0256 Data

averaged

Calculated value is the result of

an average of multiple

measurements.

0x200

0.0512

Unlocked PLL

Unlocked PLL

0x400 0.1024 Other

error

Indicates some other error

condition, usually accompanied

by a processor-specific status

description

0x800 0.2048 Other

warning

Indicates some other warning

condition, usually accompanied

by a processor-specific status

description

0x1000 0.4096 Other

info

Indicates some other informative

condition, usually accompanied

by a processor-specific status

description

0x2000 0.8192

Cumulative

result

Cumulative processing

somewhere in the chain

2-22