Status – Teledyne LeCroy WaveExpert series Automation Manual User Manual
Page 58
A
BOUT
R
ESULTS
Status
Description: Result status. Status is returned as a decimal value, but should be treated like a register with the
following bit values. See the example for decoding the Status result.
64-bit hexadecimal
int value shown in
XStreamBrowser
VT_DECIMAL
value
returned
from Status
property
StatusDescription
Description
0x1 0.0001 Invalid
result
Indicate that this Result cannot
be used for any calculation
0x2 0.0002
Data
overflow
Result contains Overflow data or
is calculated from a Result
containing Overflows.
0x4 0.0004
Data
underflow
Result contains Underflow data
or is calculated from a Result
containing Underflows
0x8 0.0008
Some values are
undefined
Result contains some values or
data that are undefined.
0x10 0.0016 Less
than
Actual value is likely to be less
than the calculated value. Mostly
used for parameter
measurements.
0x20 0.0032 Greater
than
Actual value is likely to be
greater than the calculated value.
Mostly used for parameter
measurements.
0x40
0.0064
Not a pulse
Analyzed signal is not recognized
as a pulse; mostly used for pulse
parameter measurements.
0x80 0.0128 Not
cyclic
Contains some values measured
on non-cyclic signal.
0x100 0.0256 Data
averaged
Calculated value is the result of
an average of multiple
measurements.
0x200
0.0512
Unlocked PLL
Unlocked PLL
0x400 0.1024 Other
error
Indicates some other error
condition, usually accompanied
by a processor-specific status
description
0x800 0.2048 Other
warning
Indicates some other warning
condition, usually accompanied
by a processor-specific status
description
0x1000 0.4096 Other
info
Indicates some other informative
condition, usually accompanied
by a processor-specific status
description
0x2000 0.8192
Cumulative
result
Cumulative processing
somewhere in the chain
2-22
916435 RevA