DCOM Setup

6. Click

the

Default COM Security

tab. Click the Edit Default button on the Access Permissions section of

the dialog.

Note

: Users sometimes notice the Default Properties and Default COM Security tab names sometimes are interchanged by Windows.

7. The

Access Permissions

dialog is shown. Click the Add button.

If the PC and Oscilloscope ARE in the same NT domain, this is a User Level configuration. Provide the

exact same user account (username and password) created on the PC (covered previously in this

manual).

If the PC and Oscilloscope ARE NOT in the same NT domain, this is a Share Level configuration.

Provide the exact same user account (username and password) created on the PC (covered previously

in this manual). However, the username must be preceded by the NT domain name and a backslash \.
Example

: LECROY\John.Smith.

8. Verify

the

Allow

check-box is checked on the Access Permissions dialog.

9. Click

OK

on the Access Permissions dialog and Apply, and then OK on the Properties dialog. The

Component Services

should now be showing.

10. Access the following path from Component Services: Component Services Æ Computers Æ My

Computer

Æ DCOM Config. Find and right-click on LeCroyXStreamDSO and select Properties.

A-23