beautypg.com

Chapter three – Teledyne LeCroy WaveExpert series Automation Manual User Manual

Page 180

background image

Chapter Three

Control Reference

Turns on/off TDR Normalized trace variable vertical zoom

VariableVerZoom

Bool

' Visual Basic Script
Set app = CreateObject("LeCroy.XStreamDSO")

' Turn off TDR normalized trace variable vertical zoom scale
app.TDR.TDRN.Zoom.VariableVerrZoom = "Off"

Example

Description

Turns on/off TDR Normalized trace variable horizontal zoom

VariableHorZoom

Bool

' Visual Basic Script
Set app = CreateObject("LeCroy.XStreamDSO")

' Turn on TDR normalized trace variable horizontal zoom scale
app.TDR.TDRN.Zoom.VariableHorZoom = "On"

Example

Description

TDR Normalized trace vertical zoom

VerZoom

From 0.1 to 100 step (8 digits)

Double

' Visual Basic Script
Set app = CreateObject("LeCroy.XStreamDSO")

' Assuming the normalized trace shows impedance parameter
' Change TDR normalized trace vertical scale per division to 10 ohms
app.TDR.TDRN.Zoom.VerZoom = 10

Range

Example

Description

TDR Normalized trace vertical position

VerPos

From -1.5 to 1.5 step (8 digits)

Double

' Visual Basic Script
Set app = CreateObject("LeCroy.XStreamDSO")

' Assuming the normalized trace shows S parameters
' Change TDR normalized trace center vertical position to start at -10dB
app.TDR.TDRN.Zoom.VerrPos = -10

Range

Example

Description

916435 RevA

3-113